Nox integration

Configure default values for Nox HST devices and assign them to patients.

The Recording Type is a saved setup that tells Nox devices what signals to record and how to score and report the sleep study. In Noxturnal, you can configure the Recording Type to match the needs of each study, like checking for sleep apnea or doing a full sleep test. Somnoware supports Recording Type selection for Nox A1s and T3 devices only.

In Somnoware, you can select the Recording Type at the facility level or when checking out a device.

Recording Type prerequisites

  • Contact Resmed support to enable the Recording Type in Somnoware.

  • Ensure you have SDI 9.24.9.4 or later installed.

  • Ensure the Nox device version is 7.0 or later.

  • In the Noxturnal software, ensure you have configured Recording Types. See the Noxturnal documention for more information.

Set the default values for the facility's Nox device.

  1. Click the Configuration icon ( SW_Icon_Configuration.PNG ).

  2. Next to the facility, click the Edit icon.

  3. Click the HST administrations tab.

  4. In the Supported Devices table, next to the Nox HST device, click Set Default Value.

  5. Configure the following:

    • Recording Type: Select the type of recording patients can perform with this device.

      Note

      If the Recording Type field is not available, you must enable user selection of the recording type in the Noxturnal software.

    • Manually start recording: Select this option so the patient can start the recording themselves.

    • Start recording at: Select this option for the device to start recording at the specified start time.

    • Number of nights: Enter the total number of nights for the sleep study.

    • Duration: Set the default number of hours for the device to record.

    • Start time: If you selected Start recording at, type the time you want the device to start recording the study.

  1. Set up the Nox HST device.

  2. In the Scheduler calendar, click the patient's appointment.

  3. Click Checkout Device.

  4. In the Oximeter Pin field, enter the oximeter PIN number on the device.

    Note

    If the Oximeter Pin field is not available, you must enable user selection of the recording type in the Noxturnal software.

    SW_Screenshot_OximeterPin.png
  5. Click Checkout.