Skip to main content

Somnoware

Exchange data per therapy session

Therapy data is sent after each therapy session. One ORU message is generated for each session. Each OBX segment reflects the metric recorded for that session. No normalization or averaging is applied.

The response payload includes:

  • OBX-5—the raw session-level value.

  • OBX-11—code F (Finalized).

Downstream systems may aggregate session-level data into reporting intervals.

Therapy data ORUs

For therapy data results, the following additional HL7 segments are expected:

Segment

Expected value/contents

OBR-25

Result status ("Resulted")

OBR-2 or OBR-3

Order ID

PV1-19 or PV1-51

Visit number

NTE segment(s)

User messages

Observation/Result (OBX) fields

The discrete data is sent across in HL7 OBX segments. Each OBX segment has the below components populated:

OBX field

Description

Comments

OBX-1

Set ID

OBX-2

Value type

"NM" or "ST", based on the observation value

OBX-3.1

Identifier

Discrete data identifier

OBX-3.2

Text

Discrete data description

OBX-5

Observation value

OBX-11

Observation result status

"F"- Finalized

Therapy data fields

The following therapy data fields are available:

OBX data type

(OBX-2)

OBX component

(OBX-3.1)

OBX description

(OBX-3.2)

ST

receiptTime

System time of data receipt to AirView.

ST

sessionDate

Date the data was collected.

ST

deviceID_mode

Therapy mode setting of the device on the session date.

ST

deviceID_device_serialNo

Serial number of the device used by the patient on the session date.

ST

deviceID_device_deviceType

Type of device used by the patient on the session date.

ST

deviceID_device_deviceTypeDesc

Description of the type of device used by the patient on the session date.

ST

usage_duration

Total usage in minutes of the device on the session date.

ST

usage_maskOn

Represents the number of times a user put their mask on.

ST

usage_maskOff

Represents the number of times a user took their mask off.

ST

usage_programNumber

The program this data applies to.

ST

usage_programUsage

Total usage of the program in minutes.

ST

set_press

The set pressure in cmH2O.

ST

set_startPress

The start pressure in cmH2O.

ST

set_maxPress

The maximum allowable pressure in cmH2O.

ST

set_minPress

The minimum allowable pressure in cmH2O.

ST

set_EPRType

Expiratory Pressure Relief (EPR) type.

ST

set_EPRLevel

EPR Level (EPRType must not be OFF).

ST

set_ODIThreshold

Desaturation Rule Threshold setting.

ST

set_iBREnable

Intelligent Backup Rate enabled setting.

ST

set_tgtPatientRate

Target Respiratory Rate/Target Patient Rate setting in breaths per minute.

ST

set_EPAPAutoEnable

Auto Expiratory Positive Airway Pressure (EPAP) enabled setting.

ST

set_IPAP

Inspiratory Positive Airway Pressure (IPAP) setting in cmH2O.

ST

set_maxIPAP

Maximum IPAP setting in cmH2O.

ST

set_EPAP

EPAP setting in cmH2O.

ST

set_startEPAP

Start EPAP pressure setting in cmH2O.

ST

set_autosetResponse

AutoSet Response setting.

This can be: On (Soft) or Off (Standard)

ST

set_minEPAP

Minimum EPAP setting in cmH2O.

ST

set_maxEPAP

Maximum EPAP setting in cmH2O.

ST

set_minPS

Minimum Pressure Support (PS) setting in cmH2O.

ST

set_maxPS

Maximum Pressure Support/P Control Max setting in cmH2O.

ST

set_easyBreathe

The Easy Breathe setting can be either On or Off.

ST

set_circuit

Breathing Circuit Type setting.

This can be: SingleValved, DualValved, SingleVented

ST

set_interface

Patient Interface Type setting.

This can be: Invasive, Mask, Mouthpiece

ST

set_mask

Mask Type setting on the device.

This can be: FullFace, Nasal, NasalUltra, Pillows, Pediatric, InLineVent, Infant

ST

set_pressureAssist

Pressure Assist setting in cmH2O.

Note: This is also known as Pressure Control or P Control.

ST

set_safetyTidalVolume

Safety Tidal Volume setting in mL.

ST

set_apneaVent_tidalVolume

The Tidal Volume setting in mL for apnea ventilation.

ST

set_tidalVolume

Tidal Volume setting in mL.

ST

set_ti

Inspiratory time (Ti) setting in seconds.

ST

set_apneaVent_ti

The Ti setting in seconds for apnea ventilation.

ST

set_tiMin

The Ti minimum setting in seconds.

ST

set_tiMax

The Ti maximum setting in seconds.

ST

set_riseTime

Rise Time setting in milliseconds.

ST

set_triggerSensitivity

Trigger Sensitivity setting.

This can be: VHi, Hi, Med, Lo, VLo, Off

ST

set_cycleSensitivity

Cycle Sensitivity setting.

This can be: VLo, Lo, Med, Hi, VHi

ST

set_respRate

Respiratory Rate/Backup Rate in breaths per minute.

ST

set_apneaVent_respRate

The Respiratory Rate setting in breaths per minute for apnea ventilation.

ST

set_tgtAlveolarVent

Target Alveolar Ventilation settings in liters per minute.

ST

set_triggerThreshold

Trigger Threshold setting in liters per minute.

ST

set_triggerType

Trigger Type setting.

This can be: None (off), Absolute (Flow), Relative (Pressure)

ST

set_cycleThreshold

Cycle Threshold setting in percent.

ST

set_patientType

Patient Type setting.

This can be: Adult, Pediatric

ST

set_heightCm

Height setting in centimeters.

ST

set_pressSupport

Pressure Support setting in cmH2O.

ST

set_volumeBreathOptions

Volume Breath Options setting.

This can be: TiOption (Ti), PifOption (PIF)

ST

set_flowShape

Flow Shape setting.

This can be: Constant, Dec75%, Dec50%, Dec25%

ST

set_manualBreath

Manual Breath enable setting.

ST

set_manualMagnitude

Magnitude setting in percent.

Note: Manual Breath must be enabled.

ST

set_peakFlow

Peak Flow setting in liters per minute.

Note: PIF must be selected as the Volume Breath option.

ST

set_sigh_enable

The sigh enable setting.

ST

set_sigh_alert

The sigh alert setting.

Note: The sigh enable setting must be on.

ST

set_sigh_interval

The sigh interval setting in minutes.

Note: The sigh enable setting must be on.

ST

alarmSettings_apnea_interval

The Interval/T Apnea setting in seconds for apnea alarms.

ST

set_sigh_magnitude

The sigh magnitude setting in percent.

Note: This sigh enable setting must be on.

ST

set_apneaVent_peakInspFlow

PIF setting in liters per minute for apnea ventilation.

ST

set_apneaVent_pressureControl

The Pressure Control/P Control setting in cmH2O for apnea ventilation.

ST

alarmSettings_apnea_breathType

The Breath Type/Detection setting for apnea alarms.

ST

alarmSettings_apnea_response

The Response setting for apnea alarms.

ST

clinicalMetrics_tgtlPAP_5

5% of target IPAP in cmH2O.

ST

clinicalMetrics_tgtlPAP_95

95% of target IPAP in cmH2O.

ST

clinicalMetrics_tgtlPAP_50

Median target IPAP in cmH2O.

ST

clinicalMetrics_tgtlPAP_max

Maximum target IPAP in cmH2O.

ST

clinicalMetrics_tgtEPAP_5

5% of target EPAP in cmH2O.

ST

clinicalMetrics_tgtEPAP_95

95% of target EPAP in cmH2O.

ST

clinicalMetrics_tgtEPAP_50

Median target EPAP in cmH2O.

ST

clinicalMetrics_tgtEPAP_max

Maximum target EPAP in cmH2O.

ST

clinicalMetrics_leak_5

5% of leak in liters per second.

ST

clinicalMetrics_leak_95

95% of leak in liters per second.

ST

clinicalMetrics_leak_50

Median leak in liters per second.

ST

clinicalMetrics_leak_max

Maximum leak in liters per second.

ST

clinicalMetrics_leakpermin_5

5% of leak in liters per minute.

ST

clinicalMetrics_leakpermin_95

95% of leak in liters per minute.

ST

clinicalMetrics_leakpermin_50

Median leak in liters per minute.

ST

clinicalMetrics_leakpermin_max

Maximum leak in liters per minute.

ST

clinicalMetrics_leakValved_5

5% of valved leak in percent.

ST

clinicalMetrics_leakValved_95

95% of valved leak in percent.

ST

clinicalMetrics_leakValved_50

Median valved leak in percent.

ST

clinicalMetrics_respRate_5

5% of respiratory rate in breaths per minute.

ST

clinicalMetrics_respRate_95

95% of respiratory rate in breaths per minute.

ST

clinicalMetrics_respRate_50

Median respiratory rate in breaths per minute.

ST

clinicalMetrics_respRate_max

Maximum respiratory rate in breaths per minute.

ST

clinicalMetrics_ieRatio_5

5% of inhalation:exhalation ratio.

ST

clinicalMetrics_ieRatio_95

95% of inhalation:exhalation ratio.

ST

clinicalMetrics_ieRatio_50

Median inhalation:exhalation ratio.

ST

clinicalMetrics_ieRatio_max

Maximum inhalation:exhalation ratio.

ST

clinicalMetrics_minuteVent_5

5% of minute ventilation in liters per minute.

ST

clinicalMetrics_minuteVent_95

95% of minute ventilation in liters per minute.

ST

clinicalMetrics_minuteVent_50

Median minute ventilation in liters per minute.

ST

clinicalMetrics_minuteVent_max

Maximum minute ventilation in liters per minute.

ST

clinicalMetrics_tidalVol_5

5% of tidal volume in liters.

ST

clinicalMetrics_tidalVol_95

95% of tidal volume in liters.

ST

clinicalMetrics_tidalVol_50

Median tidal volume in liters.

ST

clinicalMetrics_tidalVol_max

Maximum tidal volume in liters.

ST

clinicalMetrics_tidalVolml_5

5% of tidal volume in milliliters.

ST

clinicalMetrics_tidalVolml_95

95% of tidal volume in milliliters.

ST

clinicalMetrics_tidalVolml_50

Median tidal volume in milliliters.

ST

clinicalMetrics_tidalVolml_max

Maximum tidal volume in milliliters.

ST

clinicalMetrics_SPO2_min

Minimum SpO2.

ST

clinicalMetrics_SPO2_5

5% SpO2.

ST

clinicalMetrics_SPO2_95

95% SpO2.

ST

clinicalMetrics_SPO2_50

50% SpO2.

ST

clinicalMetrics_SPO2_minutesBelow88percent

SpO2 minutes under the 88% threshold.

ST

clinicalMetrics_SPO2_secondsBelowDynamicThreshold

SpO2 (sec) under dynamic threshold.

ST

clinicalMetrics_alveolarVentilation_5

5% of alveolar ventilation in liters per minute.

ST

clinicalMetrics_alveolarVentilation_95

95% of alveolar ventilation in liters per minute.

ST

clinicalMetrics_alveolarVentilation_50

Median alveolar ventilation in liters per minute.

ST

clinicalMetrics_alveolarVentilation_max

Maximum alveolar ventilation in liters per minute.

ST

clinicalMetrics_spontTriggBreaths

Percentage of spontaneous triggered breaths.

ST

clinicalMetrics_SpontCycledBreaths

Percentage of spontaneous cycled breaths.

ST

respEvents_AHI

Apnea-Hypopnea Index in events per hour.

ST

respEvents_HI

Hypopnea Index in events per hour.

ST

respEvents_AI

Apnea Index in events per hour.

ST

respEvents_OAI

Obstructive Apnea Index in events per hour.

ST

respEvents_CAI

Central Apnea Index in events per hour.

ST

respEvents_UAI

Unknown Apnea Index in events per hour.

ST

respEvents_ODI

Oxygen Desaturation Index.

ST

respEvents_CSR

Cheyne-Stokes Respiration in minutes.

ST

respEvents_RERA

Respiratory Effort Related Arousal in events per hour.

ST

patientInterface_humidifier

Type of humidifier used on this session date.

ST

patientInterface_heatedTube

Type of heated tube used on this session date.

ST

patientInterface_ambHumidity

Absolute ambient humidity for the session in milligrams of water per liter.

Example 7. Therapy data ORU
MSH|^~\&|SOMNOWARE|SOMNOWARE|HC|HC|20260330131022||ORU^R01|20260330131022|T|2.3
PID|1||16288888322||Sawyer^Tom^||20090325||||^^^^||^^^ruby.gold@somno.com
PV1|1||||||^^||||||||||||
ORC|RE||||||||20260330131022
OBR|1||||||20260330131022
OBX|1|ST|receiptTime^System time of data receipt to AirView||2024-03-18T09:26:17
OBX|2|ST|sessionDate^Date the data was collected||2024-03-17
OBX|3|ST|deviceID_mode^Therapy mode setting of the device on the session date||AUTOSET
OBX|4|ST|deviceID_device_serialNo^Serial number of the device used by the patient on the session date||23172404800
OBX|5|ST|deviceID_device_deviceType^Type of device used by the patient on the session date||036039
OBX|6|ST|deviceID_device_deviceTypeDesc^Description of the type of device used by the patient on the session date||AirSense 10 AutoSet
OBX|7|ST|usage_duration^Total usage in minutes of the device on the session date||392
OBX|8|ST|usage_maskOn^Represents the number of times a user put their mask on||2
OBX|9|ST|usage_maskOff^Represents the number of times a user took their mask off||2
OBX|10|ST|usage_programNumber^The program this data applies to||1
OBX|11|ST|usage_programUsage^Total usage of the program in minutes||392
OBX|12|ST|set_press^The set pressure in cmH2O||10.0
OBX|13|ST|set_startPress^The start pressure in cmH2O||4.0
OBX|14|ST|set_maxPress^The maximum allowable pressure in cmH2O||15.0
OBX|15|ST|set_minPress^The minimum allowable pressure in cmH2O||
OBX|16|ST|set_EPRType^Expiratory Pressure Relief (EPR) type||FULLTIME
OBX|17|ST|set_EPRLevel^EPR Level (EPRType must not be OFF)||THREE
OBX|18|ST|set_ODIThreshold^Desaturation Rule Threshold setting||3
OBX|19|ST|set_iBREnable^Intelligent Backup Rate enabled setting||true
OBX|20|ST|set_tgtPatientRate^Target Respiratory Rate/Target Patient Rate setting in breaths per minute||
OBX|21|ST|set_EPAPAutoEnable^Auto Expiratory Positive Airway Pressure (EPAP) enabled setting||true
OBX|22|ST|set_IPAP^Inspiratory Positive Airway Pressure (IPAP) setting in cmH2O||12.0
OBX|23|ST|set_maxIPAP^Maximum IPAP setting in cmH2O||20.0
OBX|24|ST|set_EPAP^EPAP setting in cmH2O||6.0
OBX|25|ST|set_startEPAP^Start EPAP pressure setting in cmH2O||4.0
OBX|26|ST|set_autosetResponse^AutoSet Response setting This can be On (Soft) or Off (Standard)||OFF
OBX|27|ST|set_minEPAP^Minimum EPAP setting in cmH2O||4.0
OBX|28|ST|set_maxEPAP^Maximum EPAP setting in cmH2O||10.0
OBX|29|ST|set_minPS^Minimum Pressure Support (PS) setting in cmH2O||3.0
OBX|30|ST|set_maxPS^Maximum Pressure Support/P Control Max setting in cmH2O||10.0
OBX|31|ST|set_easyBreathe^The Easy Breathe setting can be either On or Off||On
OBX|32|ST|set_circuit^Breathing Circuit Type setting This can be: SingleValved, DualValved, SingleVented||SingleVented
OBX|33|ST|set_interface^Patient Interface Type setting This can be: Invasive, Mask, Mouthpiece||Mask
OBX|34|ST|set_mask^Mask Type setting on the device This can be: FullFace, Nasal, NasalUltra, Pillows, Pediatric, InLineVent, Infant||Nasal
OBX|35|ST|set_pressureAssist^Pressure Assist setting in cmH2O Note: This is also known as Pressure Control or P Control||5.0
OBX|36|ST|set_safetyTidalVolume^Safety Tidal Volume setting in mL||400
OBX|37|ST|set_apneaVent_tidalVolume^The Tidal Volume setting in mL for apnea ventilation||401
OBX|38|ST|set_tidalVolume^Tidal Volume setting in mL||500
OBX|39|ST|set_ti^Inspiratory time (Ti) setting in seconds||1.2
OBX|40|ST|set_apneaVent_ti^The Ti setting in seconds for apnea ventilation||1.2
OBX|41|ST|set_tiMin^The Ti minimum setting in seconds||0.8
OBX|42|ST|set_tiMax^The Ti maximum setting in seconds||1.5
OBX|43|ST|set_riseTime^Rise Time setting in milliseconds||300
OBX|44|ST|set_triggerSensitivity^Trigger Sensitivity setting This can be: VHi, Hi, Med, Lo, VLo, Off||Med
OBX|45|ST|set_cycleSensitivity^Cycle Sensitivity setting This can be: VLo, Lo, Med, Hi, VHi||Med
OBX|46|ST|set_respRate^Respiratory Rate/Backup Rate in breaths per minute||14
OBX|47|ST|set_apneaVent_respRate^The Respiratory Rate setting in breaths per minute for apnea ventilation||10
OBX|48|ST|set_tgtAlveolarVent^Target Alveolar Ventilation settings in liters per minute||6.0
OBX|49|ST|set_triggerThreshold^Trigger Threshold setting in liters per minute||2.5
OBX|50|ST|set_triggerType^Trigger Type setting This can be: None (off), Absolute (Flow), Relative (Pressure)||Absolute (Flow)
OBX|51|ST|set_cycleThreshold^Cycle Threshold setting in percent||25
OBX|52|ST|set_patientType^Patient Type setting This can be: Adult, Pediatric||Adult
OBX|53|ST|set_heightCm^Height setting in centimeters||175
OBX|54|ST|set_pressSupport^Pressure Support setting in cmH2O||6.0
OBX|55|ST|set_volumeBreathOptions^Volume Breath Options setting This can be: TiOption (Ti), PifOption (PIF)||TiOption (Ti)
OBX|56|ST|set_flowShape^Flow Shape setting This can be: Constant, Dec75%, Dec50%, Dec25%||Dec50%
OBX|57|ST|set_manualBreath^Manual Breath enable setting||true
OBX|58|ST|set_manualMagnitude^Magnitude setting in percent Manual Breath must be enabled||50
OBX|59|ST|set_peakFlow^Peak Flow setting in liters per minute PIF must be selected as the Volume Breath option||40
OBX|60|ST|set_sigh_enable^The sigh enable setting||true
OBX|61|ST|set_sigh_alert^The sigh alert setting The sigh enable setting must be on||true
OBX|62|ST|set_sigh_interval^The sigh interval setting in minutes The sigh enable setting must be on||30
OBX|63|ST|alarmSettings_apnea_interval^The Interval/T Apnea setting in seconds for apnea alarms||1
OBX|64|ST|set_sigh_magnitude^The sigh magnitude setting in percent This sigh enable setting must be on||20
OBX|65|ST|set_apneaVent_peakInspFlow^PIF setting in liters per minute for apnea ventilation||35
OBX|66|ST|set_apneaVent_pressureControl^The Pressure Control/P Control setting in cmH2O for apnea ventilation||8.0
OBX|67|ST|alarmSettings_apnea_breathType^The Breath Type/Detection setting for apnea alarms||NoBreath
OBX|68|ST|alarmSettings_apnea_response^The Response setting for apnea alarms||AlarmOnly
OBX|69|ST|clinicalMetrics_tgtlPAP_5^5% of target IPAP in cmH2O||8.5
OBX|70|ST|clinicalMetrics_tgtlPAP_95^95% of target IPAP in cmH2O||12.24
OBX|71|ST|clinicalMetrics_tgtlPAP_50^Median target IPAP in cmH2O||9.0
OBX|72|ST|clinicalMetrics_tgtlPAP_max^Maximum target IPAP in cmH2O||13.8
OBX|73|ST|clinicalMetrics_tgtEPAP_5^5% of target EPAP in cmH2O||5.0
OBX|74|ST|clinicalMetrics_tgtEPAP_95^95% of target EPAP in cmH2O||8.5
OBX|75|ST|clinicalMetrics_tgtEPAP_50^Median target EPAP in cmH2O||6.5
OBX|76|ST|clinicalMetrics_tgtEPAP_max^Maximum target EPAP in cmH2O||10.0
OBX|77|ST|clinicalMetrics_leak_5^5% of leak in liters per second||0.05
OBX|78|ST|clinicalMetrics_leak_95^95% of leak in liters per second||0.48
OBX|79|ST|clinicalMetrics_leak_50^Median leak in liters per second||0.16
OBX|80|ST|clinicalMetrics_leak_max^Maximum leak in liters per second||0.78
OBX|81|ST|clinicalMetrics_leakpermin_5^5% of leak in liters per minute||3.0
OBX|82|ST|clinicalMetrics_leakpermin_95^95% of leak in liters per minute||28.8
OBX|83|ST|clinicalMetrics_leakpermin_50^Median leak in liters per minute||9.6
OBX|84|ST|clinicalMetrics_leakpermin_max^Maximum leak in liters per minute||46.8
OBX|85|ST|clinicalMetrics_leakValved_5^5% of valved leak in percent||2
OBX|86|ST|clinicalMetrics_leakValved_95^95% of valved leak in percent||10
OBX|87|ST|clinicalMetrics_leakValved_50^Median valved leak in percent||5
OBX|88|ST|clinicalMetrics_respRate_5^5% of respiratory rate in breaths per minute||10
OBX|89|ST|clinicalMetrics_respRate_95^95% of respiratory rate in breaths per minute||18
OBX|90|ST|clinicalMetrics_respRate_50^Median respiratory rate in breaths per minute||14
OBX|91|ST|clinicalMetrics_respRate_max^Maximum respiratory rate in breaths per minute||22
OBX|92|ST|clinicalMetrics_ieRatio_5^5% of inhalation:exhalation ratio||1.0
OBX|93|ST|clinicalMetrics_ieRatio_95^95% of inhalation:exhalation ratio||2.0
OBX|94|ST|clinicalMetrics_ieRatio_50^Median inhalation:exhalation ratio||1.5
OBX|95|ST|clinicalMetrics_ieRatio_max^Maximum inhalation:exhalation ratio||2.5
OBX|96|ST|clinicalMetrics_minuteVent_5^5% of minute ventilation in liters per minute||4.5
OBX|97|ST|clinicalMetrics_minuteVent_95^95% of minute ventilation in liters per minute||8.2
OBX|98|ST|clinicalMetrics_minuteVent_50^Median minute ventilation in liters per minute||6.0
OBX|99|ST|clinicalMetrics_minuteVent_max^Maximum minute ventilation in liters per minute||10.0
OBX|100|ST|clinicalMetrics_tidalVol_5^5% of tidal volume in liters||0.3
OBX|101|ST|clinicalMetrics_tidalVol_95^95% of tidal volume in liters||0.6
OBX|102|ST|clinicalMetrics_tidalVol_50^Median tidal volume in liters||0.45
OBX|103|ST|clinicalMetrics_tidalVol_max^Maximum tidal volume in liters||0.8
OBX|104|ST|clinicalMetrics_tidalVolml_5^5% of tidal volume in milliliters||300
OBX|105|ST|clinicalMetrics_tidalVolml_95^95% of tidal volume in milliliters||600
OBX|106|ST|clinicalMetrics_tidalVolml_50^Median tidal volume in milliliters||450
OBX|107|ST|clinicalMetrics_tidalVolml_max^Maximum tidal volume in milliliters||800
OBX|108|ST|clinicalMetrics_SPO2_min^Minimum SpO2||88
OBX|109|ST|clinicalMetrics_SPO2_5^5% SpO2||90
OBX|110|ST|clinicalMetrics_SPO2_95^95% SpO2||98
OBX|111|ST|clinicalMetrics_SPO2_50^50% SpO2||95
OBX|112|ST|clinicalMetrics_SPO2_minutesBelow88percent^SpO2 minutes under the 88% threshold||5
OBX|113|ST|clinicalMetrics_SPO2_secondsBelowDynamicThreshold^SpO2 (sec) under dynamic threshold||120
OBX|114|ST|clinicalMetrics_alveolarVentilation_5^5% of alveolar ventilation in liters per minute||3.5
OBX|115|ST|clinicalMetrics_alveolarVentilation_95^95% of alveolar ventilation in liters per minute||7.5
OBX|116|ST|clinicalMetrics_alveolarVentilation_50^Median alveolar ventilation in liters per minute||5.5
OBX|117|ST|clinicalMetrics_alveolarVentilation_max^Maximum alveolar ventilation in liters per minute||9.0
OBX|118|ST|clinicalMetrics_spontTriggBreaths^Percentage of spontaneous triggered breaths||85
OBX|119|ST|clinicalMetrics_SpontCycledBreaths^Percentage of spontaneous cycled breaths||80
OBX|120|ST|respEvents_AHI^Apnea-Hypopnea Index in events per hour||2.1
OBX|121|ST|respEvents_HI^Hypopnea Index in events per hour||0.6
OBX|122|ST|respEvents_AI^Apnea Index in events per hour||1.5
OBX|123|ST|respEvents_OAI^Obstructive Apnea Index in events per hour||1.5
OBX|124|ST|respEvents_CAI^Central Apnea Index in events per hour||0.0
OBX|125|ST|respEvents_UAI^Unknown Apnea Index in events per hour||0.0
OBX|126|ST|respEvents_ODI^Oxygen Desaturation Index||1.8
OBX|127|ST|respEvents_CSR^Cheyne-Stokes Respiration in minutes||0
OBX|128|ST|respEvents_RERA^Respiratory Effort Related Arousal in events per hour||0.0
OBX|129|ST|patientInterface_humidifier^Type of humidifier used on this session date||internal
OBX|130|ST|patientInterface_heatedTube^Type of heated tube used on this session date||None
OBX|131|ST|patientInterface_ambHumidity^Absolute ambient humidity for the session in milligrams of water per liter||7


Report request ORMs

This order provides the information to get an unsolicited PDF report from AirView for a patient. You can request Compliance and Therapy reports.

For report request orders, the following additional HL7 segments are expected:

Segment

Expected value\Contents

Required

OBX-3 (code) and OBX-5

The report is delivered in English (en-us) unless a language is specified (see Supported PDF languages).

OBX-3 (code) and OBX-5

If periodType = USERDEFINED, this field is required.

If periodType = COMPLIANCE, this field is ignored.

Yes

OBX-3 (code) and OBX-5

If periodType = USERDEFINED, this field is required.

If periodType = COMPLIANCE, this field is ignored.

Yes

OBX-3 (code) and OBX-5

Must be one of the following two values:

  • COMPLIANCE

  • USERDEFINED

Yes

Supported PDF languages

Language value

Language of generated report

de

German

en

UK English

en-us

US English

es

Spanish

fr

French

fr-ca

French Canadian

ja

Japanese

pt

Portuguese (Portugal)

sv

Swedish

it

Italian

Example 8. Report request ORM
MSH|^~\&|HCP1|HCP1|Resmed|Resmed|20250313092149||ORM^O01|21020683|T|2.5|||||||||

PID|1||257^^^MRN||||||||||||||||||||||||||||||||||||

PV1|1|U||||||||||||||||||||||||||||||||||||||||||||||||||

ORC|NW|21028683|||||||20250313|||||||||||||||||||||

OBR|1|21020683||RES104^THERAPY REPORT|3|20250313||||||||||||||||||||||||||||||||||||||||||

OBX|1|DT|startDate^Report Start Date:^RESLQL||2025-11-30||||||F||||||||

OBX|2|DT|endDate^Report End Date:^RESLQL||2025-12-10||||||F||||||||

OBX|3|TX|periodType^Period Type:^RESLQL||USERDEFINED||||||F||||||||